2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 181
CLEAN- 4.6.1 BOTH PDIA AND ADIA ARE GOOD
a) Return the dialyzer connector to the shunt and close the door!
b) Press ‘CONFIRM’ then
return to HEAT DISINFECT!
c) To isolate between four (4) possible bad valves read part d’s two-part description BEFORE performing
part e!
d) Return to debug screen 4.
PDIA
NORMALLY
behaves as follows:
PART 1: When Valve #43 is closed*
PDIA
cycles between 0.0 (for 3 to 5 seconds) to more than 0.1
(for about one second). It is abnormal for
PDIA
to cycle to 6.0 or more OR remain 0.0 for more than
seven (7) seconds!
*
Valve #43 cycles between closed for forty five (45) seconds and open for sixteen (16) seconds
PART 2: When Valve #43 is open*
PDIA
stays between 1.0 and 5.0. When Valve #43 closes
PDIA
resumes cycling. If (and ONLY if)
PDIA
REMAINS between 1.0 and 5.0 for MORE THAN ninety (90)
(page 182).
e) WITHOUT LOOKING AWAY, watch
PDIA
AND for “Flow Recirc Error 1”
for five (5) minutes OR until if
“Flow Recirc Error 1” appears. Proceed according to one of the TWO (2) possible scenarios below:
1)
IF (and ONLY if) “Flow Recirc Error 1” NEVER occurs:
Whatever was causing the Flow Error
is not exhibiting yet! Watch
PDIA
until either a Flow Error occurs OR Heat Disinfect finishes.
2) IF “Flow Recirc Error 1” occurs AND also one of THREE (3) scenarios below:
Scenario #1: IF (and ONLY if) PDIA was EVER = 6.0 OR more:
A restriction at Valve #24 OR
Valve #25 is indicated. See the Figures below to locate these valves!
Scenario #2: IF (and ONLY if) PDIA EVER stayed 0.0 for longer than seven (7) seconds:
A
restriction at Valve #29 is indicated! See the left Figure below to locate Valve #29!
Scenario #3: IF PDIA NEVER = 6.0 OR more AND NEVER 0.0 more than seven (7) seconds:
To isolate Valve #43, proceed to
page 183
, procedure number
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...