Yes
Between 4.0 and 6.0 volts!
(page 629).
No
IS NOT between 4.0 and 6.0 volts!
The Power Logic Board is bad (card cage, first circuit board
from the left).
P- 1.0.2 PIN 3 BETWEEN 4.0 AND 6.0 VOLTS / ISOLATE POWER ON/OFF CIRCUIT
a) Spread the card cage side panels open then gently drop the front panel down.
b)
Per the Figure below
, TWO (2) checks:
CHECK #1:
ENSURE the twenty-four (24) pin Front Panel Data Cable is plugged in properly between
the motherboard and the Front Panel Interface board. If not this may be the problem!
CHECK #2:
ENSURE the sixteen (16) pin, Switch Matrix Cable is plugged in securely, pin for pin, at
the Front Panel Interface board. If not this may be the problem!
c) Measure at the 16 pin Switch Matrix Cable’s, from
pin 4
(4
th
pin from the REAR of the machine).
Between 4.0 and 6.0 volts DC?
Yes
Between 4.0 and 6.0 volts! Proceed to
page 631
, procedure number
No IS NOT between 4.0 and 6.0 volts! See procedure number
(page 630).
Figure 121 – Front Panel
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...