2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 487
PHT- 4.6.0 UF INTEGRITY TEST PASSED
a) Return the UF Pump’s tubing.
b) Was the machine originally pulled for service because of reported “
On Line PHT Failed”?
Yes
On Line PHT Failed! See procedure number
(page 487).
page 488
PHT- 4.7.0 ON LINE PHT FAILED / DIAGNOSTIC VALVE LEAK TESTS / VALVE #24 AND #25
ONLY two (2) valves will be tested, Valves #24 and #25:
a)
Place the machine into Service Mode → Diagnostics → Valve Leak Test.
b) Allow the
[Test Status]
data box = “Ready”.
c) Select the screen’s
[Valve Number]
data box.
d) K machine use the up/down arrow keys until
[Valve Number]
= “Valve #24”; K
2
machine use the
+/- keys until
[Valve Number]
= “Valve #24”.
e) Press ‘CONFIRM’.
f) Allow “Prep” and “Testing” to complete then read the
[TEST STATUS]
box, TWO (2) possible scenarios:
1) IF “Failed”:
REPEAT
the test on the leaking valve. If (and ONLY if) it fails
AGAIN
TWO (2)
possibilities:
1)
Replace the Actuator-Test Board with a known good then repeat the Valve Leak
Test to test on the valve. If (and ONLY if) the valve fails again:
2)
Leaking valve! To locate the
valve refer to
2) IF
“
Passed”:
REPEAT
the Valve Leak Test on “Valve #25”. Do BOTH valves #24 and #25 pass?
Yes
BOTH Valves #24 AND #25 pass! See procedure number
(page 488).
No A valve fails! TWO (2) possibilities:
1)
Replace the Actuator-Test Board with a known good
then repeat the Valve Leak Test to test it. If (and ONLY if) the test fails again:
2)
Bad valve!
To locate the valve refer to
LEFT BLANK INTENTIONALLY
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...