2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 116
F- 15.0.0 LOADING PRESSURE LOW / ISOLATE DEAERATION MOTOR RATE (DEAP)
A) Turn the Heater Switch OFF for now!
B) Call debug screen 1 to see
DEAP
(upper right). TWO (2) possible scenarios:
1)
IF (and ONLY if) DEAP = 75 OR LESS!
(page 117).
2)
IF DEAP = 76 OR more:
See parts a THROUGH h below:
a) Enter
Service Mode → Calibrate Hydraulics → Deaeration Pressure.
DO NOT follow the
screen instructions! See part b instead!
b) Press ‘CONFIRM’. The screen’s
[Pump Rate]
window turns yellow.
c) Select the
[Pump Rate]
window to turn it bright yellow.
d) Set
[Pump Rate]
to “210”!
e)
Press ‘CONFIRM’ TWICE to save the calibration!
f) Turn the machine OFF!
g) Turn the machine on and return to Dialysis Program (“Select Program”
→ ‘Dialysis’ →
‘CONFIRM’).
h) Call debug screen 1. Is
DEAP
now = 45?
Yes
DEAP
(page 117).
No
DEAP
NOT = 45! CAREFULLY repeat the Deaeration Pressure calibration from part a.
but, if (and ONLY if) after repeating
DEAP
STILL does not = 45 THREE (3) possible
bad components:
1)
Actuator-Test Board OR;
2)
Functional Board EEPROM (IC2) OR;
3)
Functional Board.
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...