![Fresenius Medical Care 2008K2 Скачать руководство пользователя страница 340](http://html.mh-extra.com/html/fresenius-medical-care/2008k2/2008k2_troubleshooting-manual_2331635340.webp)
2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 336
CO- 1.0.10 VERIFY ACID SETTINGS
a) Select the upper left ‘Conc’ button to call the ACID Sub Screen (Figure below). This screen indicates
the ‘Selected Acid’s’
Ionic Profile
.
ACID Sub Screen
b) Does the label on the connected ACID match the Sub Screen’s
Ionic Profile
(Yes or No)?
Yes
The label = the screen’s Ionic Profile. See parts a AND b below:
a) Per its label, ENSURE the bicarb is the same ‘Type’ as the acid.
page 337
).
No Referring to NOTES 1 through 4 below, THREE (3) possibilities:
1)
The connected acid may
NOT be the ‘Selected Acid’;
2)
You made an error and the levels are correct;
3)
The connected
acid may have been entered incorrectly in Service mode.
NOTE 1:
If a Granuflo
®
acid is selected the screen’s
Acetate (Ac)
value should be ½ the value on the
ACID’S label. EXAMPLE: 4.0 on the screen = 8.0 on the label. If Granuflo
®
IS NOT selected
the screen’s
Ac
value should = exactly the value on the label!
NOTE 2:
Sub Screen values are rounded! EXAMPLE: 2.30 on the screen may = 2.25 on the label.
NOTE 3:
Na+
= Sodium. If the sodium level on the label is 110 or less it indicates sodium contributed
by the acid alone. If more than 110 it indicates the TOTAL or FINAL Sodium.
NOTE 4:
K+
= Potassium; Ca
++
= Calcium; Mg
++
= Magnesium;
Ac
= Acetate
;
Citrate
is NOT
ALWAYS present;
Dex
= Dextrose (mg/dL
*
).
*
If the label states Dextrose in mmol/L THEN 5.5 mmol/L = 100 mg/dL;11 mmol/L = 200
mg/dL
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...