2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 143
NW- 1.0.2 “NO WATER” CHECKS
a) ENSURE no leaks, ESPECIALLY from Heat
Exchanger #77 (Figure right).
b) Call debug screen 0.
c) WITHOUT LOOKING AWAY, watch
Valve #41’s ‘dot’ (Figure right) for ninety
(90) seconds. TWO (2) possible
scenarios:
1)
IF (and ONLY if) Valve #41’s ‘dot’ remains
white OR cycles between white and blue:
See
(page 143).
2) IF Valve #41’s ‘dot’ is ALWAYS blue (NEVER white)!
ENSURING the water or portable RO is on
AND the station’s incoming water valve is open, if the ‘dot’ still REMAINS blue, proceed to
page
, procedure number
NW- 1.0.3 INTERMITTENT “NO WATER” ALARM
NOTE:
If (and ONLY if) in a CLEANING / DISINFECTION PROGRAM AND a “No Water” alarm is coming
and going the PROGRAM will complete but takes longer than normal. If you can live with this, especially if
using a portable RO, the machine is okay for now.
WITHOUT LOOKING AWAY, watch for four (4) FULL minutes OR until if a “No Water” alarm EVER
reoccurs?
Yes
“No Water” reoccurs! See procedure number
(page 143).
No Does NOT reoccur! Do NOT continue!
NW- 1.0.4 “NO WATER” ALARM REOCCURED
Leaving the machine in whatever Program it is CURRENTLY in, TWO (2) possible scenarios:
1)
IF (and ONLY if) in HEAT DISINFECT Program:
(page 144).
2)
ALL OTHER Programs including Dialysis:
Proceed to
page 146,
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...