2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 262
T- 1.8.5 INCOMING WATER MORE THAN 20.0 ° C / TROUBLESHOOT LOW TEMPERATURE
a)
To avoid damage turn the machine OFF!
b) Open the card cage and reseat the sensor, Power Logic and Functional Boards!
(page 262).
T- 1.8.6 REATTEMPT TEMP CONTROL CALIBRATION
a) Return to
Service Mode → Calibrate Sensors → Temp Control.
b) ‘Sharply’ press ‘CONFIRM’ TWICE to turn the screen’s
[TEMP DAC]
data box pale yellow/white.
c) Select
[TEMP DAC]
, it turns bright yellow.
d) Set
[TEMP DAC]
to “170”.
e) Press ‘CONFIRM’.
[TEMP DAC]
MUST return to pale yellow / white i.e. NOT gray.
f) ENSURING flow through the external indicator, allow up to six (6) minutes OR until if the external meter
reaches 35.5° C or more?
Yes
35.5° C or more! Proceed to
page 306
, procedure number
No Remains less than 35.5° C! Perform parts a AND b below:
a)
To prevent damage, turn the machine OFF!
b)
Read before performing!
Swap in the listed components (see COMPONENT LIST below),
one at a time, with known good, then in between, repeat procedure number
(page
262) to test each new component until the meter’s temperature is more than 35.5° C!
COMPONENT LIST: 1)
NTC #2
1
;
2)
Power Logic Board;
3)
Sensor Board
2,3
,
4)
Actuator-Test Board;
5)
Functional Board
2,3
6)
Sensor Board cable;
7)
Power Control
board (located inside the power supply);
8
) Distribution board;
9)
Motherboard.
1
To LOCATE NTC #2, refer to Figure 42 (page 218)
2
To LOCATE the board, refer to
(page 9)
3
To prevent “Cond Offset Failure”, place the machine into
T and C Mode
(refer to
OPERATING MODES
(page
Error! Bookmark not defined.
))
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...