2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 265
T- 2.0.2 T5 LESS THAN 20.0 VOLTS
a)
CAUTION!
The next measurement will be from the Power Logic board’s twenty-pin X2 cable, at
pin
7
.
Pin 9, next door, is connected to 24 volts DC and
touching pins 7 and 9 together WILL
DAMAGE several boards!
To prevent this, TWO (2) precautions:
1) Use the protected red meter lead!
2) Figure below, place the lead horizontally on pin 7
b) Measure at
PIN 7
, (BOTTOM row, four pins from the REAR of the machine). 4.0 volts DC or MORE?
Yes
Pin 7 is 4.0 volts OR MORE! See procedure number
(page 265).
No
Pin 7 is LESS THAN 4.0 volts! Proceed to
page 268
, procedure number
T- 2.0.3 PIN 7 IS 4.0 VOLTS OR MORE
All card cage signals appear to be good!
a)
Set your meter AC voltage (~ V, V
AC
)!
b) NTC #2 remains unplugged till instructed!
Parts c and d next page