2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 574
WD- 1.0.4 CHECK FOR POWER CONTROL BOARD ‘SHORT’
Turn the machine ON. TWO (2) possible scenarios:
1)
IF (and ONLY if) “WD: Fail Long Pulse” OR “WD: Fail Short Pulse” does NOT reoccur:
Turn the
machine off and on several times, allowing 40 seconds in between, If “WD: Fail Long Pulse” OR “WD:
Fail Short Pulse” does NOT reoccur the new IC4 fixed the problem!
2)
IF “WD: Fail Long Pulse” OR “WD: Fail Short Pulse” REOCCURS:
See procedure number
(page 574).
WD- 1.0.5 ISOLATE POWER SUPPLY / CARD CAGE
a)
IMPORTANT!
ENSURE the
“WD: Fail Long Pulse” OR “WD: Fail Short Pulse”
banner is up
BEFORE continuing to part b.
b) Per the Figure below, measure again at the TEST Connector’s,
pin 8
. TWO (2) possible scenarios:
1)
IF (and ONLY if) more than 3 volts DC:
Assuming all above procedures were performed
correctly, there may be a 24 volt ‘short circuit’ inside the power supply or the card cage. See
procedure number
(page 575).
2) IF less than 3 volts DC:
The symptom changed from what originally brought you here. It appears
the new IC4 is working. Return to (ABOVE) procedure number
(page 571) to re-measure
voltage and proceed as directed.
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...