2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 274
T- 2.2.0 PIN 6 LESS THAN 9.0 VOLTS DC / ISOLATE POWER LOGIC BOARD
a) Spread the card cage side panels open then gently drop the front panel down!
b) Per the Figure below, at the bottom of the
Sensor Board
, a row of pins extend horizontally, turn 90
degrees downward, and connect to the Sensor Board’s motherboard connector. This is the "C” row,
numbered 1 through 32 (from front-to-rear).
c) Expecting 3.6 volts or LESS, measure
at pin C1 (very FRONT pin). THREE (3) possible scenarios below:
1) IF 9.0 volts or more:
See procedure number T- 2.2.1 (page 275).
2) IF between 3.7 and 9.0 volts:
ENSURING no accidental contact with adjacent pin C2 see
procedure number T- 2.2.1 (page 275).
3) IF 3.6 volts or LESS:
See parts a THROUGH d below:
a)
Turn the machine OFF!
b)
TWO (2) possible bad components (see Component List below). Swap in one at a time, with
known good, and continue to part c to test the component.
c)
Place the machine into Dialysis or Heat Disinfect.
d)
Measure again at the Power Logic Board’s X2 cable,
pin 6
. When 9.0 volts OR MORE is
measured the last component swapped in is bad. If still LESS THAN than 9.0 volts repeat
parts a through d.
COMPONENT LIST:
1)
Power Logic Board (see Figure 4A and NOTE A, page 9);
2)
Motherboard.
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...