2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 395
CO- 12.0.00 OLC SELF TEST PASSED / CHECK BLOOD PUMP / ADDITIONAL TROUBLESHOOTING
OLC Self-Test PASSED! A conductivity-related OLC problem is not indicated HOWEVER, unstable Blood
Pump rate may also cause OLC Tests to cancel. See parts a THROUGH c below:
a)
IMPORTANT!
Reset ALL alarms. The machine MUST remain alarm free!
b) Turn the Blood Pump on to 400 ml/min and allow two (2) FULL minutes without changing
its rate again! If (and ONLY if) the Pump is stable
1
, from debug screen 5,
QbS
(BOTTOM
window, Figure right) will settle to and remain between 0 and 3.
1
Resetting multiple blood alarms cause Blood Pump rate instability
(OLC
cancel code
= “3”). In this event the problem may be related to the patient’s
access.
c) ENSURE ALL treatment parameters (see below) were entered PRIOR to treatment initiation. Changes
made just prior to OR while an OLC Test is running cause instability AND OLC Test cancelation.
● Remaining Time of Dialysis (RTD)
•
UF Goal UF Time
•
Fluid Removed (UF Removed)
•
Sodium Variation (SVS) should be
activated just after UF is turned on
● Dialysate Flow Rate
•
Blood Flow Rate
•
Temperature
•
Concentrate (bath)
•
Na+ and Bicarbonate settings
d) For additional OLC diagnostics call debug screen 6 and refer to
e) For even more OLC diagnostics call debug screen 5 and refer to
(page 396).
Table A - Debug Screen 6 OLC Values
Parameter
Description
Value
Max QBS
Maximum
QbS
(screen 5)
during the OLC test
Less than 20 indicates stable Blood Pump rate. If
more than 20 the OLC test cancels. See also
QB
Cancel
and
OLC Cancels 1 - 6
.
QD Begin
Dialysate Flow Rate at
beginning of OLC Test
Valid ONLY if patient is on machine
8 = 800 ml/min; 5 = 500 ml/min, etc
QD Cancel
Dialysate Flow if OLC test
canceled
Valid ONLY if patient is on machine
5 = 500 ml/min, 3 = 300 ml/min, etc
OLC Cancels 1- 6
a
a
NOTE:
If the machine
is turned off
OLC
Cancels 1 – 6
codes
default to 0!
0 =
OLC test not yet performed
1
= Unable to reach or stabilize @15.5 mS
2
= Unable to reach or stabilize @13.5 mS
3
= Blood pump rate changed or was turned off or
multiple blood alarms occurred
4
= Dialysate Flow was unstable or turned off or
multiple TMP alarm resets or Filling Programs
or Conductivity or Temperature alarms
occurred
5
= OLC was turned OFF
6
= Dialysate Flow was turned off
7
= OLC test started
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...