2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 319
CO- 1.0.0 ISOLATE DIALYSATE SETTINGS
A) At the bottom of the screen, select the ‘Dialysate’ tab.
B) Figure below, ENSURE the ‘Selected Acid’ matches the acid connected to the machine.
C) Figure below, ENSURE the Acid ‘Type’ matches the acid label.
Figure 55 – Dialysate Screen
D) Set [
Base Na+
] to 137 and press ‘CONFIRM’.
E) Set [
Bicarbonate
] to 33 and press ‘CONFIRM’.
F) If the ‘Selected Acid’, [
Base Na+
] OR [
Bicarbonate
] was changed, allow five (5) minutes BEFORE
continuing to part G.
G) Based on [
Conductivity
] now, THREE (3) possible scenarios:
1)
IF (and ONLY if) LESS THAN 13.0 mS:
(page 320).
2)
IF (and ONLY if) STABLE between 13.0 and 14.3:
Proceed to
page 335
, procedure number
3)
ALL OTHERS:
Proceed to
page 325
, procedure number