2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 424
FIL- 2.4.4 FILACT = 1
a) Figure right, swap in a known good Air Sensor Connector.
b)
Figure below
, be
SURE
to plug the new Air Sensor’s
Connector into its distribution board position PROPERLY! If CBE
equipped, using a flashlight, ENSURE the connector
covers ALL
CBE board pins, especially the TOP pin
!
c) Figure right, connect the new Connector’s wires together.
d) Return to Dialysis Program (“Select Prog
ram →‘Dialysis’
→ ‘CONFIRM’).
d) Allow thirty (30) seconds BEFORE continuing to part e.
e) Call debug screen 1. WITHOUT LOOKING AWAY, watch
FILACT
for two (2) minutes or until if it EVER = 1. TWO
(2) possible scenarios:
1)
IF (and ONLY if) FILACT = 0 ALWAYS:
Problem solved! The previous Air Sensor Connector is
bad!
2)
IF FILACT EVER = 1:
Leaving the wires connected together until
FILACT
= 0 ALWAYS, perform
parts A through F below:
A)
To avoid damage turn the machine OFF!
B) SIX (6) possible bad components (see Component List below). One at a time, with the
machine off, swap in each with known good then in between continue with part C through F to
test each new component.
Component List:
1)
CBE board;
2)
Sensor Board ;
3)
Sensor Board cable;
4)
Functional
Board
;
5
) Distribution board;
6)
Motherboard
C) If the time comes to swap in the Sensor or Functional board, to prevent “Cond Offset”
Failure”, place the machine into
T and C Mode
(refer to
OPERATING MODES
(page
Error!
Bookmark not defined.
));
D)
Turn the machine on and return to Dialysis Program (“Select Program →‘Dialysis’ →
‘CONFIRM’).
E) Allow thirty (30) seconds BEFORE continuing to part F!
F) Call debug screen 1. Watch
FILACT
for two (2) minutes. If EVER = 1 return to part A) and
swap in the next component in the list. When
FILACT
remains = 0 the last component
swapped in is the problem.
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...