2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 437
FIL- 6.0.0 PDIAL DOES NOT AGREE WITH THE METER
a) To prevent wetting the meter’s transducer protector clamp the ‘to meter’ tubing segment!
b) Turn Dialysate Flow ON (Dialysate Flow on/off lamp STOPS blinking).
c) From the Home screen, set [Dialysate Flow] to 500 ml/min and press ‘CONFIRM’.
d) Enter
Service Mode → Calibrate Sensors →
Dialysate Pressure. The screen says
“1. Connect a
pressure gauge in line…
”. The “gauge” (meter) is ALREADY connected!
e) ENSURE the external flow indicator’s ‘bob’ is moving up and down.
f) Press ‘CONFIRM’. The screen says
“3. Press the [Dialysate Flow on/off] key…”.
g) Press the Dialysate Flow on/off key (the Flow on/off lamp blinks).
h) ENSURE the external flow indicator’s ‘bob’ is NOT moving (i.e. Flow is off).
i)
IMPORTANT!
Remove the clamp from the external meter’s tubing segment!
j)
Using the syringe, adjust pressure until the external meter reads 0 +/- 2 mmHg.
(page 437).
FIL- 6.1.0 CREATE NEGATIVE PRESSURE
a) Press ‘CONFIRM’. The screen now says
“6. Pressurize until dialysate pressure reads -250
mmHg….”
.
b) PULL on the syringe plunger to achieve negative (-)250 +/- 5 mmHg on the external meter.
c) Clamp the Four-Way’s syringe tubing
to
keep this pressure!
d) ENSURING the external meter = -250 +/- 5 mmHg, press ‘CONFIRM’.
e) Figure right, TWO (2) possible scenarios based on
if an “Operator Error” banner occurs:
1)
IF (and ONLY if) “Operator Error” occurred!
(page 438).
2) IF “Operator Error” DID NOT occur!
See parts a AND b below:
a) Press ‘CONFIRM’ twice to save the calibration.
b) Proceed to
page 439
, procedure number
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...