2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 481
h) When “Get Pos TMP” MOMENTARILY appears
ATMP
should drop to b300 and +500.
i)
When “Pos Flow Off” first appears a twenty (20) second stabilization period begins.
ATMP
should settle
b180 and +350.
NOTE:
If
FILACT
EVER = 1 during “Pos Flow Off” NOTE this!
NOTE:
If
ATMP
does not remain b180 and +350 “Fail Pos Stabilize” will be listed in the Test
& Options screen possibly indicating a large external leak OR
FILACT
went to ‘1’.
j)
Assuming
ATMP
remained b180 and +350 during the stabilization period a 30 second HOLD
period begins. “Pos Flow Off” remains up AND
ATM
P should not change more than +/- 30
NOTE:
If
ATMP
changes more than +/- 30 “Fail Pos Flow Off” will be listed in the Test & Options
screen possibly indicating a small external leak.
k) When the “Testing” banner appears IGNORE
ATMP
and
FILACT
. Do BOTH Pressure Tests pass?
Yes
BOTH pass! Proceed to
page 484
, procedure number
No “Pressure Test(s) Failed”! See procedure number
(page 481).
PHT- 3.1.2 PRESSURE TEST(S) FAILED AGAIN
Select the Test & Options tab. FOUR (4) possible scenarios 1) or 2) or 3) or 4) below:
1)
IF (and ONLY if) “Fail Remove Air” is listed:
Proceed to
page 409
2) IF (and ONLY if) “Fail Neg Stabilize” is listed.
TWO (2) possible scenarios i) or ii) below:
i)
IF FILACT EVER ‘1’ (even if only once) during the repeat 30 second STABILIZATION
“Neg Flow On” period:
Proceed to
page 409
ii)
IF FILACT ALWAYS ‘0’ during the repeat 30 second “Neg Flow On” STABILIZATION
period:
Proceed to
page 440
SECTION 9 – TMP (PRESSURE) PROBLEMS
3) IF (and ONLY if) “Fail Neg Flow On” is listed:
THREE (3) possible scenarios i) or ii) or iii) below:
i)
IF ATMP increased to more than -440:
(page 482).
ii)
IF FILACT EVER ‘1’ (even if only once) during the repeat 30 second HOLD “Neg Flow On”
period:
Proceed to
page 409
SECTION 8 – FILLING PROGRAM PROBLEMS
iii)
IF FILACT ALWAYS ‘0’ during the repeat 30 second HOLD “Neg Flow On” period:
Proceed to
page 409
SECTION 8 – FILLING PROGRAM PROBLEMS
4) ALL Positive Test Failure listings i.e. “Fail Get Pos TMP”, “Fail Pos Stabilize”, “Fail Pos Flow
Off”:
Negative Pressure Test failures have priority over Positive Test failures. If any Negative Pressure
test failed see above Scenarios 2 or 3. If a Negative test did not fail TWO (2) possible scenarios i) or ii)
next page:
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...