VE- 1.0.2 ISOLATE VERR
a) Call debug screen 0.
Figure 37 – Debug
Screen 0
b)
Figure above
, watch the
Balancing Chamber Valve
‘dots’, for one (1) minute. THREE (3)
possible scenarios 1) or 2) or 3) below:
1)
IF (and ONLY if) ALL EIGHT are REMAINING white:
See procedure number
2) IF (and ONLY if) four (4) are REMAINING blue AND four (4) are REMAINING white:
Proceed to
page 200
, procedure number
3)
IF ALL EIGHT are CYCLING between white and blue:
Proceed to
page 199
, procedure
number
VE- 2.0.0 ALL EIGHT BALANCING CHAMBER VALVES REMAINING WHITE
One of the valve circuits whose ‘dots’ are blue is the problem!
These procedures isolate between them:
a)
Open the shunt door and leave it open till instructed!!
b) Allow up to thirty (30) seconds. Do the Balancing Chamber Valve ‘dots’ begin to cycle between white
and blue?
Yes
The Balancing Chamber Valve ‘dots’ cycle! Valve #25 OR its blue wire harness is bad! To
LOCATE Valve #25 refer to
(page 193).
No
All Balancing Chamber Valve ‘dots’ REMAIN white! Leaving the shunt door open, TWO (2)
possible scenarios next page:
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...