2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 113
LESS THAN 800 ml collected continued:
a) ENSURE no tubing restrictions at the DiaSafe
®
filter.
b) Figure below, open Dialysate Filter #73’s housing and ENSURE the filter inside is clean!
c) Per the Figure below, touch Valve #25’s black solenoid. Is it warm?
Yes
The solenoid is warm! TWO (2) possible bad components:
1)
Bad Actuator-Test
Board OR;
2)
Bad Valve #25
No The solenoid is NOT warm!
NOTE
ONLY
VALVE #25
will be checked and proceed
to
page 192
Figure 23 – Filter #73 / Valve #25
F- 14.0.5 MORE THAN 800 ml COLLECTED / OUT OF BYPASS CIRCUIT IS OKAY
a) Turn the machine OFF!
b) Turn the machine on and return to Dialysis Program with [Dialysate Flow] set to 800 ml/min.
c) Open the shunt door then, if present, reset a TMP alarm up to twice before continuing.
d) Call debug screen 6. WITHOUT LOOKING AWAY, watch
BC Switch
(middle
column) for five (5) miinutes or until it EVER = 897 or more. THREE (3) possible
scenarios 1) or 2) or 3) below:
1)
IF BC SWITCH ALWAYS = 897 or more:
Return to
page 30
, procedure number
2)
IF BC SWITCH = 897 or more intermittently OR if a previously observed Flow Error was
occuring very intermittently:
Return to
page 40,
3)
IF BC SWITCH NEVER = 897 or more:
The Troubleshooting Guide cannot locate a Flow
problem at this time.
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...