2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 54
F- 6.1.4 PDIA STABLE BETWEEN 1.0 AND 8.5
a) From the Home screen, ENSURE [Dialysate Flow] is set to 800 ml/min.
b) [
Conductivity
] MUST BE more than 11.0 mS indicating the Acid Pump is drawing!
c) Is
[
Conductivity
] more than 14.6 mS?
Yes
[
Conductivity
] more than 14.6 mS! Proceed to
page 62
No
[
Conductivity
] less than 14.6 mS! See parts A THROUGH E below:
A) Call debug screen 0. Figure right, the
Balancing Chamber valve ‘dots’ MUST be
cycling, possibly every nine (9) seconds.
B) If Valve #24’s ‘dot’ is blue ENSURE the external
flow indicator’s ‘bob’ is rising and falling.
C) Valve #41’s ‘dot’ should cycle between white
and blue almost every time the Balancing
Chamber valves cycle indicating drain flow.
D)
Figures below
, ENSURING TMP is STABLE b200 and negative 400 (stable =
DOES NOT change more than 40 per minute), watch
ACFS
for two (2) minutes. TWO (2)
possible scenarios below:
Scenario #1:
ACFS NORMAL behavior:
Mostly remains between 2.0 and 5.0 but may
occasionally drop below 1.0 but for
NO MORE
than ten (10) seconds. If
(and ONLY if)
ACFS
is behaving normally continue to part D.
Scenario #2: ACFS ABNORMAL behavior:
If EVER 5.9 or more OR remains lower than 1.0
for LONGER THAN ten (10) seconds proceed to
page 89
, procedure number
D) Call debug screen 6 to see
BC Switch
(middle column). It normally
changes slightly (+/- 10) but if EVER more than 400 OR less than 203
indicates a flow problem. In any event see part E next page.
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...