2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 285
Yes Drain Flow continues!
Read this step before performing it!
The previous Valve #30
is probably good. The Actuator-Test Board is bad. After replacing the board return to
(ABOVE) procedure number
(page 283) to see if this fixes the temperature
problem!
No Drain flow stopped! Return to (ABOVE) procedure number
(page 283) to see if
this fixes the temperature problem.
T- 3.0.4 DRAIN FLOW STOPPED
/ PREPARE TO MEASURE ‘ACTUAL’ TEMPERATURE
a)
IMPORTANT!
Press ‘Escape’ then ‘CONFIRM’ twice to call the “Select Program” screen.
b) Remove the dialyzer lines from the shunt and attach them to a calibrated Temperature (°C) meter
*.
*
DO NOT USE a Mesa
®
NEO-1
as it CANNOT measure temperature more than 40°C! The NEO-2
and most other meters can be used! Refer to the meter’s Operator’s Manual!
c) Figure right, under the
shunt door locate the
three (3) spring loaded
switches
d) See procedure number
(page 285).
T- 3.0.5 HEAT DISINFECT / MEASURE ‘ACTUAL’ TEMPERATURE
This procedure simulates the dialyzer connectors in the shunt door to allow the Heat Disinfect Program to
start!
a) With your right hand PUSH IN AND HOLD
ALL THREE
(3) switches.
b) With your left hand, place the machine in
HEAT DISINFECT
.
c) WITHOUT closing the door RELEASE the buttons.
d) The “Cover is Open” banner appears! A “Flow Recirc Error” may occur but MUST go away within forty
five (45) seconds!
e) ENSURING the flow indicator’s ‘bob’ is moving up and down, see procedure number T- 3.0.6 (page
286).
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...