2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 659
DVF- 4.2.2
BOTH MEASUREMENTS BETWEEN 40 AND 100 Ω
a) Turn the machine ON.
b) Return to Dialysis Program (
“Select Program” → ‘Dialysis’ → ‘CONFIRM’)!
c) If (and ONLY if) a “Dial Valve Failure” OR “Act Byp Valve
Fail” alarm banner reoccurs see procedure number
(page 659). If neither occurs allow five (5)
minutes BEFORE continuing to part d!
d) Call debug screen 1. Is
VERR
(right column, bottom) = 0?
Yes
VERR
= 0! See (ABOVE) procedure number
(
page 656
).
No
VERR
= 1 OR more! Call debug screen 0. Ignoring
Flow Error
, for one (1) minute, watch the
2
nd
window down
,
Valve Error!
Ignoring a ‘blink to 1’ that lasts less than one (1) second,
does
Valve Error
EVER = 1 for LONGER THAN two (2) seconds?
Yes
Valve Error
= 1 for LONGER THAN two (2) seconds! See procedure number
No
Valve Error
= 0 OR ‘blinks to 1’ for less than one second! Proceed to
page
TROUBLESHOOTING VALVE ERRORS IN DIALYSIS PROGRAM
DVF- 4.3.0 VALVE ERROR = 1 LONGER THAN TWO SECONDS
Assuming all procedures were performed correctly, THREE (3) possible bad components (see COMPONENT
LIST below). Swap in each, one at a time, and in between return to Dialysis Program to test the new
component until debug screen 1’s
VERR
remains = 0.
COMPONENT LIST:
1)
Actuator-Test Board;
2)
Functional Board
1
;
3)
Distribution board.
1
To prevent “Cond Offset Failure”, place the machine into
T and C Mode
(refer to
OPERATING MODES
,
page
Error! Bookmark not defined.
)).
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...