2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 359
CO- 6.5.0 “COND LESS THAN 10 or 15” BANNER OCCURING
a) Recheck Loading Pressure (gauge in Rinse port). Unless there is a masked “No Water” or a Flow Error
alarm pressure cycles, to between NEVER less than 11 psi and a PEAK of about 25 psi. Do NOT
continue UNLESS pressure is cycling approximately EVERY THREE SECONDS!
b) ENSURING the shunt door is CLOSED, if the external flow indicator’s ‘bob’ is NOT moving up and
down there may be an unannounced “No Water” alarm or a Flow Error. Do NOT continue UNLESS the
‘bob’ is moving up and down!
c) Look through the clear acid and bicarb inlet tubing to ENSURE the pumps are drawing into the
machine with NO AIR! If a problem is located and fixed allow six (6) FULL minutes before continuing to
part d!
d) Does the screen’s
[Conductivity]
box go to between 13.0 and 14.0 at steps 9 or 10 OR between 15.0
and 15.9 at steps 8 or 10?
Yes
(page 359).
CO- 6.5.4 [CONDUCTIVITY] BOX NOT GOOD (2)
a)
IMPORTANT!
Swap acid and bicarbonate with a machine that has good conductivity.
b) Allow six (6) FULL minutes.
c) Does the
[Conductivity]
box stabilize between 13.0 and 14.0 mS at steps 9 or 10 OR between 15.0
and 15.9 mS at steps 8 or 10?
Yes
(page 359).
No Return to (ABOVE) procedure number
CO- 6.5.6 [CONDUCTIVITY] BOX STABILIZES
Does the “COND LESS THAN 10 or 15…” alarm go away?
Yes
Alarm goes away! Return to (ABOVE) procedure number
(page 358).
No
a) ‘Escape’ the calibration
b) Return to (ABOVE) procedure number
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...