2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 476
PHT- 1.5.0 INITIAL PHT CHECKS
a) Call the Home screen. With the Level Detector’s P
ven
port vacant,
(Figure right)
Venous Pressure
MUST = 0 mmHg!
b) With [Dialysate Flow] set to 500 ml/min or more:
•
[Temperature] MUST be stable
*
between 35.0 and 39.0° C!
•
[Conductivity] MUST be stable
*
between 13.0 and 14.3 mS!
*
Stable = NOT changing more than 0.2 per minute
c) Using a flashlight, ENSURE no air bubbles MOVING into the machine
through the acid and bicarb inlet tubing!
d) Install a ‘dummy venous chamber’ in the Level Detector module.
e) RESET ALL alarms!
f) If screen says “Standby For Test” OR the tests are running
(screen says “Test:…”) allow them to finish BEFORE continuing to procedure number
PHT- 1.5.1 INITIAL PHT CHECKS CONTINUED
IMPORTANT!
If a “No Water” or Flow Error alarm EVER occur
address them first!
a) Call debug screen 0. Valve #24’s ‘dot’ (Figure right) MUST be
BLUE BEFORE continuing to part b!
b) From the Home screen, set [Dialysate Flow] to 500 ml/min and
press ‘CONFIRM’.
c) Small external leaks cause BIG problems! Using a flashlight, FOUR (4) checks. BE THROUGH!
Check #1:
Figure right, the ENTIRE length of the dialyzer
lines ESPECIALLY the dialyzer quick connectors!
Check #2:
The hydraulic compartment! Puddles on
the compartment floor may indicate an
active leak!
Check #3:
Figure right, from the DiaSafe
®
filter housing and tubing
connections!
Check #4:
ENSURE the external flow indicator’s (sight tube) fittings are
tight!
(page 477).
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...