2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 441
N) Figure right, if necessary adjust the Conductivity
Limits until the ‘Actual’ Conductivity is CENTERED
between them.
O)
IMPORTANT!
Press ‘CONFIRM’.
P) Allow up to three (3) minutes for the Conductivity
window to turn white indicating no Conductivity
alarms!
Q)
IMPORTANT! CLOSE THE SHUNT DOOR!
R) Call debug screen 2. To ENSURE the shunt door is
indicated closed
CVRCLS
(2
nd
column from left) = 1!
S) Call debug screen 0.
T) Allow Valve #24’s ‘dot’ (Figure right) to turn blue BEFORE
CONTINUING to part U indicating no Temp or Cond alarms!
U) Is the flow indicator’s ‘bob’ move at least ¼ way up in the sight
tube?
Yes
‘Bob’ moving! See procedure number
(page 442).
No ‘Bob’ NOT moving! See parts a THROUGH c below:
a) From the Home screen, ENSURE [Dialysate Flow] remains at 500 ml/min!
b) From debug screen 0. ENSURE Valve #24’s ‘dot’ is
REMAINING
BLUE!
c) Is the external flow indicator’s ‘bob’ rising and falling now?
Yes
‘Bob’ moving! See procedure number
(page 442).
No ‘Bob’ NOT moving! ENSURE Valve #24’s ‘dot’ is REMAINING BLUE! If (and
ONLY if) ‘bob’ is still not moving, proceed to
page 20
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...