2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 192
TROUBLESHOOTING A VALVE
A)
To avoid damage, turn the machine OFF!
B)
There is NO NEED to check all 17 valves!
Check ONLY the valve as
NOTED in the procedure that brought you here!!
C) See procedure number TSV- 1.0.0 below:
TSV- 1.0.0 CHECK DISTRIBUTION BOARD CONNECTIONS
a)
below
, remove the distribution board cover.
b) TWO (2) checks to ENSURE:
Check #1:
The ACTUATOR CABLE is plugged in SECURELY AND is not damaged!
Check #2:
The
NOTED
valve is plugged PROPERLY into its position.
If not, this most likely is the
problem!
Figure 34 –
Distribution Board
c) Before continuing to procedure number
(page 193) OR
(page 193) to LOCATE the
NOTED
valve.