2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 513
BL- 2.0.0 BLOOD LEAK ALARM PRESENT (CHECK FOR AIR)
IMPORTANT!
Air in dialysate cause false blood leak alarms. Per the Figure below,
using a flashlight
,
watch for air bubbles through the Blood Leak Sensor’s TO AND FROM tubing for one (1) minute each?
Yes
Air seen! See procedure number
(page 514).
No air seen! See procedure number
page 516
).
Figure 89 – Hydraulics Side View
LEFT BLANK INTENTIONALLY