2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 544
OVER- 3.0.1 ISOLATE INLET PRESSURE REGULATOR (Slow Overflow)
This procedure checks the Regulator #61’s ability to maintain pressure over time.
a)
IMPORTANT!
Leaving the water on, turn the machine
OFF
!
b) NOTE gauge pressure for use later. It MUST NOT be more than 22 psi!
c) Allow up to fifteen (15) minutes! Does pressure increase more than 2 psi above what was noted in part b
(Yes or No)?
Yes
Pressure increased more than 2 psi! TWO (2) possibilities:
1)
Incoming water pressure
more than 105 psi OR;
2)
Bad Regulator #61. There is a Rebuild kit for Regulator #61 (P/N
190934).
No
Pressure did NOT increase! See procedure number
(page 544).
OVER- 4.0.0 ISOLATE INLET WATER VALVE #41 (Slow Overflow)
a) ENSURE the machine is off AND the water is on!
b)
Per the Figure below
, at the top of the hydrochamber, locate the clear tubing that vertically enters
the Air Gap tower. This tubing is from Inlet Water Valve #41.
Figure 101 – Valve #41 Check
c)
DO NOT touch the Vent Tubing
but pull the tubing off the Air Gap Tower.
Part d next page
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...