2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 563
LEAKING- 4.0.1 CANNOT PUSH ALL 60 ML THROUGH THE SAMPLE PORT
a) Remove the syringe from the Sample Port and fill it with water.
b) Per the Figure right, remove Valve #30’s INPUT tubing!
c) Attach a tubing segment to the syringe that will fit snug onto the
syringe AND Valve #30’s INPUT nozzle.
d) Attach the syringe to Valve #30. Push the plunger
AS HARD
AS YOU CAN
! Can you push ALL 60 ml through Valve #30
within SIX (6) seconds?
Yes
All 60 ml can be pushed out! UF Check Valve #64* may be restricted. *To LOCATE Check
Valve #64 refer to
(page 562).
No Cannot push all water out! See procedure number
(page 563).
LEAKING- 4.0.2 CANNOT PUSH THROUGH VALVE #30 / ISOLATE VALVE #30
a) Per the Figure right, remove the
BOTTOM
‘To DRAIN’ tubing from the rear of the
machine!
b) Fill the syringe with 60 ml of water and
reattach it to Valve #30’s INPUT nozzle.
c) Push
AS HARD AS YOU CAN
. Can
you now push 60 ml through Valve #30
within SIX (6) seconds? You should feel
very little resistance!
Yes All 60 ml can be pushed! The drain
tubing OR the “Quick Connector” (if
present) is restricted!
No Cannot push all water out!
NOTE
ONLY
VALVE #30
will
be checked and proceed to
page 192
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...