2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 658
2)
IF (and ONLY if) BOTH valves between 40 and 100
Ω:
3) IF ONE OR BOTH is less than 40 OR more than 100
Ω:
Replace the valve AND its blue wiring
harness.
DVF- 4.2.0 BOT
H VALVES BETWEEN 40 AND 100 Ω / ISOLATE TOTAL CIRCUIT RESISTANCE (Ώ)
a)
IMPORTANT!
Return BOTH valves connectors to their PROPER distribution board positions!
b) To avoid pulling cables loose, GENTLY open the card cage!
c)
IMPORTANT! Connect the volt meter’s black lead to chassis ground
(see
(page 4)).
d) Figure below, TWO (2) measurements below, from the solder (rear) side of the Actuator-Test board’s P2
connector. Both should be between 40 and 100
Ω
!
Measurement #1: Pin #46
(TOP row, 3rd
pin
from the screen) = Valve #43
Measurement #2: Pin #29
(BOTTOM row, 11th
pin
from the screen) = Valve #24
e) TWO (2) possible scenarios:
1)
IF (and ONLY if) BOTH are between 40 and 100
Ω:
659).
2) IF ONE OR BOTH LESS THAN 40
Ω OR MORE THAN 100 Ω:
Perform parts a THROUGH c
below:
a) ENSURE the machine was OFF AND valves #24 and #43 were returned to the distribution board
properly! If NOT, return to (ABOVE) procedure number
(page 658).
b) ENSURE both measurements were made at the correct Actuator-Test Board P2 pins!
c) THREE (3) possible bad components:
1)
Bad Actuator-Test Board OR;
2)
Bad actuator cable
OR;
3)
Bad distribution board.
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...