2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 313
T- 7.6.6 ADJUST ‘SLOW FLOW’ MONITOR REFERENCE
a) ‘Sharply’ press ‘CONFIRM’ to turn
[Monitor Reference]
pale yellow / white!
b) Adjust
[Monitor Reference]
to
match the meter’s maximum (i.e. between 36.9 and 37.1° C).
c) CAREFUL HERE! Sharply ‘CONFIRM’ TWICE to turn
[Monitor Reference]
GRAY!
d) Does the GRAY
[Monitor Reference]
data box read between 36.4 and 37.6° C?
Yes
The GRAY data box is between 36.4 and 37.6° C! This completes the calibration HOWEVER,
if you started out troubleshooting a “TEMP OVER 95 DEGREES” alarm return to procedure
number
page 247
). If not, return to Dialysis Program.
No
The GRAY data box IS NOT between 36.4 and 37.8! Swap in the listed components (see
COMPONENT LIST below), one at a time, with known good, in between returning to (ABOVE)
procedure number
(page 305) to test each new component.
COMPONENT LIST: 1)
NTC #3
1
;
2)
Functional Board.
1
To LOCATE NTC #3 refer to
(page 218)
LEFT BLANK INTENTIONALLY
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...