2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 175
CLEAN- 4.0.0 GOOD FLOW PUMP PRESSURE / ISOLATE THE CLEANING PROGRAM
a) Leave the gauge installed
ENSURE
its tubing is NOT kinked (restricting flow)!
b)
Open the shunt door and LEAVE IT OPEN until instructed!!
c) Leaving the machine in whatever Program it is CURRENTLY in:
1)
Rinse? OR
2)
Heat Disinfect?
1)
IF (and ONLY if) in RINSE:
Proceed to
page 184
, procedure number CLEAN- 5.0.0.
2) IF in HEAT DISINFECT:
See parts a AND b below:
a) Press ‘Escape’ to call the Heat
Disinfect screen (Figure right).
b) Ignoring the
Remaining Time
window, does the VERY TOP
window,
Remaining Prerinse
Time
, = 0:00 min:sec?
Yes
Remaining Prerinse Time
= 0:00!
See procedure number
(page 175).
No
Remaining Prerinse Time
does
NOT = 0:00! Proceed to
page
, procedure number
CLEAN- 4.2.0 REMAINING PRERINSE TIME = 0:00
a) Call debug screen 0. If the debug screens do not appear
press ‘Escape’ then call screen 0.
b) When
Remaining Prerinse Time
= 0:00 Valve #29’s ‘dot’
(Figure right) REMAINS blue. WITHOUT LOOKING AWAY,
watch it for twenty (20) FULL seconds.
c) TWO (2) possible scenarios:
1)
IF (and ONLY if) Valve #29’s ‘dot’ is ALWAYS blue:
(page 176).
2) IF Valve #29’s ‘dot’ is EVER = white:
Proceed to
page 184
, procedure number CLEAN- 5.0.0.
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...