2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 334
CO- 1.0.8.5 AFTER RINSE CONDUCTIVITY = 10.0 MS
a) Select the [Dialysate Flow] window.
b) Set [Dialysate Flow] to 800 ml/min and press ‘CONFIRM’.
c) Allow six (6) minutes BEFORE continuing to procedure number CO- 2.0.0 (
page 337
).
LEFT BLANK INTENTIONALLY