2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 564
LEAKING- 5.0.0 ‘TO DRAIN’ PATH OKAY / ISOLATE VALVE #24
a) A 1000 ml graduated cylinder is required!
b) ENSURE the external flow indicator’s ‘bob’ is at the bottom of the sight tube!
c) Flow from the shunt door and the flow through the indicator will be checked. Parts d through f MUST be
performed QUICKLY so read them BEFORE performing!
d) Place the machine into
RINSE
!
e) Remove the RED dialyzer connector from the shunt door but
DO NOT
shut the door!
f) Figure right, strong flow, more than 300 ml every thirty (30)
seconds?
Yes More than 300 ml! BEFORE returning the connector
to the shunt and while rinse is still running ENSURE
the external indicator’s ‘bob’ is moving, at least ½ way
up, through the sight tube. See procedure number
(page 564).
No Not more than 300 ml per 30 seconds! Turn the machine OFF then perform parts a THROUGH c
below:
a) ENSURE no tubing restrictions to and from the DiaSafe
®
filter including inside its housing!
b) Replace the DiaSafe
®
filter with a primed one from another machine
*
!
*
NOTE:
Using a new unprimed filter may cause error!
c) Repeat (ABOVE) procedure number
(page 564). If still not more than 300 ml
per thirty (30) seconds!
NOTE
ONLY
VALVE #24
will be checked and proceed to
page
LEAKING 5.1.0 VALVE #24 OKAY / ISOLATE VALVE #25
a) Press the ‘Escape’ key then ‘CONFIRM’ twice to call the “Select Program” screen.
b)
IMPORTANT!
Return the dialyzer connector to the shunt and close the door!
c) Parts d and e MUST be performed quickly! Read them BEFORE performing!
d)
IMPORTANT!
Return to RINSE Program!
e) Is the external flow indicator’s ‘bob’ rising at least ½ way?
Yes
‘Bob’ moving! Valve #25 is okay! See procedure number
(page 565).
No
‘Bob’ NOT moving! TWO (2) possible scenarios 1) or 2) next page:
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...