2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 325
CO- 1.0.4 ISOLATE ACID PUMP CONTROL
a)
Open the shunt door and LEAVE IT OPEN till instructed!
b) Call debug screen 0. Figure right, WITHOUT LOOKING AWAY, watch
Acid
versus
AAcid
for one (1) minute. Are they ALWAYS /- 3 of each other?
Yes
AAcid
=
Acid!
No
AAcid
is NOT remaining within 3 of
Acid!
There is a problem with Acid Pump control! Close the
shunt door THEN proceed to
page 397
SECTION 6 - CONCENTRATE PUMP ERRORS
...
CO- 1.0.4.2 AACID = ACID / ISOLATE BICARB PUMP CONTROL
Figure right, watch
Bic
versus
ABic
for one (1) minute. Are they remaining always
/- 3 of each other?
Yes
ABic
=
Bic!
(page 325).
No.
ABic
is NOT remaining /- 3 of
Bic!
Figure right, watch the Bicarb
(BIC) pump’s ‘dot’ for THREE (3) minutes. It should REMAIN blue. TWO
(2) possible scenarios:
1)
IF (and ONLY if) the ‘dot’ is EVER white:
(page 374).
2) IF the ‘dot’ REMAINS BLUE:
There is a problem with Bicarb Pump control. Close the shunt
door THEN
proceed to
page 397
SECTION 6 - CONCENTRATE PUMP ERRORS
CO- 1.0.5 ABIC = BIC / ISOLATE FILLING PROGRAM
Filling Programs (“Air” in dialysate) may mask Loading Pressure problems and Flow Errors.
a) From debug screen 0, Figure right, watch the text box
above Chamber #69 for thirty (30) seconds for it to
EVER
say “Air”, even just once!
b) TWO (2) possible scenarios:
1)
IF it EVER says “Air” even just once:
(page 326).
2) IF (AND ONLY if) it ALWAYS says “No Air”!
Proceed to
page 328
, procedure number
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...