2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 596
P- A.4.0 ISOLATE ‘HEATER CIRCUIT’
a)
CAUTION!
Turn the machine OFF! Electrocution hazard if NOT turned off!
b) Figure right, at the Distribution board’s Heater
Connector, TWO (2) checks:
Check #1: TOP:
From right-to-left, wires MUST be
brown (hot), Blue (neutral),
Green/yellow stripped (ground).
Check #2: BOTTOM:
From right-to-left, wires
MUST be brown, blue,
green/yellow
stripped (
US
Heater
) OR black, white,
green
(
German Heater
).
c) From the BOTTOM of the Heater Connector DISCONNECT: If a
US Heater
the brown wire OR if a
German Heater
the black wire.
d)
IMPORTANT! Turn the Heater Breaker Switch
on!
e) If possible turn the machine on (fan running).
f) If possible place the machine in Heat Disinfect then allow up to thirty (30) minutes.
g) Does the GFI ‘blow’ again?
Yes
GFI ‘blows’ again! FOUR (4) possible bad components:
1)
Bad Fan OR;
2)
Bad Power
Control board (inside the power supply) OR;
3)
Bad power supply (or power cord / plug) OR;
4
) Bad distribution board.
No Does NOT ‘blow’ again! Perform parts a THROUGH c below:
a) Replace the heater. To LOCATE the heater refer to
(page 129).
b) After replacing the heater turn the Heater Switch on.
c) From Dialysis Program OR Heat Disinfect if the GFI ‘blows’ again the fan or the power
supply may be bad.
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...