2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 103
F- 10.0.0 BAD ACFS SIGNAL
a) Figure right, ENSURE the CFS #10
Pressure Transducer’s connector is
plugged PROPERLY into the
distribution board.
F- 10.1.0 CFS PLUGGED IN PROPERLY
a) Call debug screen 10. If all procedures were performed correctly
PDIA
(middle column)
= between 3.5 and 6.5.
b) Unplug the CFS’s connector from distribution board position X10, “CFS”.
c) Using a flashlight, check inside the vacant “X10” position.
If corrosion or damaged pins are located
this may be the problem!
d) Figure right, plug the Dialysate
Pressure Sensor’s (#9) connector
into the CFS Pressure Sensor’s.
“P-DIAL”
→ “CFS”.
e)
ACFS
(lower right)
should go to between 4.0
and 6.0?
Yes
ACFS
between 4.0 and 6.0! The CFS Pressure Sensor #10* is bad. ^To LOCATE the
CFS Pressure Sensor #10 refer to
(page 92).
No
ACFS
is less than 4.0 OR more than 6.0! Leaving PDIA in CFS for now, FIVE (5) possible bad
components:
1)
Actuator-Test Board
a
OR;
2)
Sensor Board
b
OR;
3)
Sensor Board cable OR;
4)
Distribution board OR;
5)
Motherboard.
a
A)
With the power off, swap in a known good Actuator-Test
1
Board.
B) IMPORTANT!
Return Dialysis Program
(“Select Program” → ‘Dialysis’ →
‘
CONFIRM’).
C)
If
ACFS
is now
between 4.0 and 6.0 the previous Actuator-Test Board is bad.
1
To LOCATE the board refer to Figure 4A (page 9).
b
A)
With the power off, swap in a known good Sensor Board
2
;
B)
Place the machine into
T
and C Mode
(refer to
OPERATING MODES
(page
Error! Bookmark not defined.
);
C)
IMPORTANT!
Return to Dialysis Program!;
D)
If
ACFS
is now = between 4.0 and 6.0 the
previous Sensor Board is bad.
2
To LOCATE the Sensor board see Figure 4A (page 9).
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...