2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 593
SECTION 25 - TROUBLESHOOTING POWER DISTRIBUTION
A)
BEFORE continuing to part B, perform
B) If troubleshooting the screen ‘remaining black’, if air is felt from the Power
Supply vents (Figure right) = fan running = MACHINE IS ON!
C) Figure right, the Main Power Switch does not turn the machine on but must be
on to turn the machine on! It MUST REMAIN ON for all procedures!
D) Proceed according to the OBSERVED symptom ONLY! Listed below are
eleven (11) possible symptoms:
1) IF ‘blows’ the GFI i.e. wall breaker:
A)
Turn off ALL OTHER machines that are connected to GFI
circuit;
B)
Reset the GFI;
C)
Plug a known good machine into it;
D)
If possible, place it into Heat
Disinfect THEN allow up to thirty (30) minutes for the GFI to possibly ‘blow’ again. TWO (2) possible
scenarios
i)
or
ii)
below:
i)
IF (and ONLY if) the GFI ‘blows’ again:
Either the GFI is not rated for the current drawn
when a machine is in Heat Disinfect OR it may be bad.
ii) IF the GFI DOES NOT ‘blow’ again:
There may have been too many machines plugged into
the circuit. Plug in the suspected malfunctioning machine and (if possible) place it into Heat
Disinfect. If (and ONLY if) the GFI ‘blows’ again see procedure number
(page 595).
2)
IF “24V Low” OR “24V High” OR “WD: 24V Rcvr Err Short” OR “WD: 24V Rcvr Err Long”:
Proceed to
page 597
, procedure number
3) IF “5V Low” OR “5V High“:
Proceed to
page 597
, procedure number
4) IF “12V POWER FAIL“:
Proceed to
page 597
, procedure number
5) IF the fan is running BUT, after fifty (50) seconds, the screen REMAINS BLACK:
Proceed to
page
, procedure number
6) IF the machine turns on (fan runs) BUT THEN turns itself off:
Proceed to
page 647
, procedure
Scenarios 7 through 11 next page
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...