BL- 6.0.0 BLOOD LEAK ALARM WILL RESET?
Press and release the ‘RESET’ key up to three (3) times in twenty (20) second intervals. TWO (2) possible
scenarios below:
1)
IF the Blood Leak alarm CAN be cleared:
The most likely cause of a false blood leak alarm is “air in
dialysate” possibly caused when priming a dialyzer or moving the lines between the shunt and the
dialyzer. If SURE there is no air flowing through the Blood Leak Sensor see procedure number
(page 520).
2)
IF the Blood Leak alarm cannot be cleared:
(page 520).
BL- 6.1.0 CLEAN / CALIBRATE BLOOD LEAK SENSOR #8
a) Turn the machine OFF!
b) Using ESD precautions push down HARD on ALL circuit boards in the card cage. After reseating
ENSURE they are locked in!
c) ENSURE the Sensor and Actuator board ribbon cables are plugged in SECURELY
d)
IMPORTANT!
Perform ACID CLEAN followed by BLEACH DISINFECT to ensure the Blood Leak
Detector’s glass tube is clean of bicarb residue!
WARNING!
Calibrating Blood Leak without first performing a Bleach Disinfect may lead to
unnecessary parts replacement and / or cause an inaccurate Blood Leak calibration!
(page 520).
BL- 6.2.0 CALIBRATE BLOOD LEAK SENSOR #8
a)
Place the machine into Service mode → Calibrate Sensors → Blood Leak Detector.
b) Press ‘CONFIRM’ to start the calibration then see part c.
c) NORMALLY the calibration, one at a time, takes the screen’s
Leak
then
Dimness
windows to
approximately 5.00 and should stabilize within no more than FOUR (4) minutes.
d) Does the screen advance to step #5 “
Calibration complete, press the [CONFIRM] key to save
” within
FOUR (4) minutes (Yes or No)?
Yes
Calibration completes! Press ‘CONFIRM’ twice to save the calibration. Problem solved for
now
HOWEVER,
if a false Blood Leak alarm reoccurs in the near future swap the listed
components (see
Component List
below) one at a time with known good and repeat
(page 520) until a Blood Leak does NOT reoccur.
No
Turn the machine off and swap the following components (see
Component List
below), one at
time, with known good, and in between attempting the Blood Leak calibration until it calibrates
normally.
Component List: 1)
Blood Leak Sensor #8
1
;
2)
Actuator-Test Board
2
;
3)
Sensor Board
2
;
4)
Sensor
Board cable;
5)
Functional Board
2
;
6)
Distribution board.
1
NOTE:
To LOCATE the Blood Leak Sensor see
(page 513)
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...