![Fresenius Medical Care 2008K2 Скачать руководство пользователя страница 448](http://html.mh-extra.com/html/fresenius-medical-care/2008k2/2008k2_troubleshooting-manual_2331635448.webp)
2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 444
TMP- 1.0.4 ISOLATE LOADING PRESSURE
a)
ENSURE
the Loading Pressure gauge (yellow connector) reads 0 psi before inserting it!
b)
SLAM
*
the gauge into the red Acetate/Acid rinse port.
*
OR ELSE pressure will not be read correctly!
c) Loading Pressure is ‘OKAY’ if it achieves a PEAK of between 23 and 25 psi AND never cycles below
11 psi.
d) Watch the gauge for one (1) FULL minute. FOUR (4) possible scenarios 1) or 2) or 3) or 4):
1)
IF (and ONLY if) pressure REMAINS ‘OKAY’:
(page 445).
2) IF (and ONLY if) peaking to MORE THAN 26 psi:
Turn Valve #65’s nut (Figure below)
counterclockwise (outward). If pressure CAN be adjusted to between 23 and 26 psi, see procedure
number
(page 445). If (and ONLY if) it CANNOT Valve #65 may be bad.
3)
IF (and ONLY if) pressure is NEVER less than 11 psi AND peaks to between 18 and 22:
Adjust Valve #65’s nut (Figure below). Clockwise increases pressure! If PEAK pressure CAN be
adjusted to between 23 and 25 psi proceed to
page 446
, procedure number
. If (and
ONLY if) it CANNOT proceed to
page 27
procedure number
F- 1.0.8
.
4)
ALL other scenarios:
ENSURING the gauge was SLAMMED into the Rinse Port proceed to
to
page 27
procedure number
F- 1.0.8
.
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...