2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 23
b) Figure right, TWO (2) checks:
Check #1:
At the top of the distribution board,
ENSURE the Sensor Cable is
plugged in SECURELY!
Check #2:
ENSURE a resistor plug has NOT
been prematurely plugged into the
5
th
position from the left!
c) Turn the machine ON and return to Dialysis
Program (“Select Program”
→ ‘Dialysis’ →
‘CONFIRM’)!
d) Call debug screen 0. Based on Chamber #69’s
text box now, TWO (2) possible scenarios:
1) IF it says “Air” always:
Return to ABOVE
2) IF it continues to EVER say “No Air”:
Filling
Program is not working!
See parts a AND b below:
a) Before performing part b NOTE this page number THEN perform
(page 6).
b) FIVE (5) possible bad components:
1)
CBE board
1
;
2)
Sensor Board
2
;
3)
Functional Board
3
;
4)
Distribution board;
5)
Motherboard.
1
A)
Swap in a known good CBE board;
B)
With the Air Sensor’s connector
REMAINING
UNPLUGGED
return to Dialysis Program (“Select Program” → ‘Dialysis’ → ’CONFIRM’);
C)
From debug screen 0, if Chamber #69’s text box now says “Air” the previous CBE board
is bad.
2
A)
With the machine off, swap in a known good Sensor Board;
B)
Place the machine into
T
and C Mode
(refer to
OPERATING MODES
(page
Error! Bookmark not defined.
);
C)
With the Air Sensor’s connector
REMAINING UNPLUGGED
,
return to Dialysis Program;
D)
From debug screen 0, if Chamber #69’s text box now says “Air” the previous Sensor Board
is bad.
3
A)
With the machine off, swap in a known good Functional Board;
B)
Pace the new
Functional Board into
T and C Mode;
C)
With the Air Sensor’s connector
REMAINING
UNPLUGGED
, return to Dialysis Program;
D)
From debug screen 0, if Chamber #69’s text
box now says “Air” the previous Functional Board is bad.
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...