2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 22
1) IF says “Air” always:
See procedure number
(page 22).
2) IF says “No Air” OR toggles between “Air” and “No Air”:
See
parts a AND b below:
a) Inside the distribution board, ENSURE the Air
Sensor’s female connector is UNPLUGGED
from
5
th
position
FROM THE LEFT!
b) TWO (2) possible scenarios:
1)
IF (and ONLY if) Chamber #69 now says
“Air” always:
See procedure number
(page 22).
2)
IF Chamber #69 continues to say “No Air” or toggles between “Air” and “No Air”:
Proceed to
page 22
, procedure number
F- 1.0.2.11 CHAMBER #69 SAYS “AIR” ALWAYS
a) Call debug screen 1.
b) Allow two (2) minutes OR until if
FILACT
(middle column) = 1?
Yes
FILACT
= 1! Proceed to
page 24
, procedure number
No
AFTER two (2) minutes
FILACT
REMAINS = 0!
Look at
VERR
(lower right).
TWO (2) possible scenarios:
1)
IF (and ONLY if) VERR = 0:
See procedure number
(page 22).
2)
IF VERR = 1 or more:
Proceed to
page 30
, procedure number
F- 1.0.3 CHAMBER #69 SAYS “NO AIR” / TROUBLESHOOT FILLING PROGRAM FAILURE
a)
Turn the machine OFF!
Parts b through f next page
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...