2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 151
NW- 3.0.0 ISOLATE FLOAT (#5) SIGNAL
a)
IMPORTANT!
TURN THE WATER OFF!
b) Figure right, place the
274 Ω
resistor plug, from
the
TWO-RESISTOR SET
, into the float’s
distribution board position, “FLOAT-SW”. This
simulates a ‘closed’ float switch!
c) Call debug screen 0. WITHOUT LOOKING
AWAY, watch VALVE #41’s dot’ for two (2)
minutes. TWO (2) possible scenarios:
1)
IF (and ONLY if) Valve #41’s ‘dot’ is BLUE always:
Valve #41 should be open. Leaving the
plug installed, see procedure number
(page 152).
2) IF Valve #41’s ‘dot’ is or EVER turns WHITE:
See parts a AND b below:
a) ENSURE the
274
Ω
plug, from the
TWO-RESISTOR SET
is PROPERLY placed at
distribution board position “FLOAT-SW”! If not, repeat procedure number
(page 151)
from part b.
b) Leaving the plug installed, swap the following components in (Component List below) one at a
time and, in between, test each new component until Valve #41’s ‘dot’ remains blue indicating
the last component swapped in is the problem.
Component List: 1)
Actuator-Test Board
1
;
2)
Sensor Board
1,2
;
3)
Sensor Board cable;
4)
Distribution board;
5)
Motherboard. .
1
To LOCATE the boards refer to
(page 9).
2
To prevent “Cond Offset Failure”, place the machine into
T and C Mode
(refer to
OPERATING MODES
, page
Error! Bookmark not defined.
)
LEFT BLANK INTENTIONALLY
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...