2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 35
a) Momentarily plug JUST the acid (red) connector into its rinse port to call “Select Program”!
b) RETURN the acid connector to concentrate!
c) Select the ‘Dialysis’ button but
DO NOT
press ’CONFIRM’
TILL INSTRUCTED
!
d) Call debug screen 0. If parts a through c were performed correctly the TOP balancing chamber
valves ‘dots’, #31 through #34, are REMAINING BLUE!
e) Remove the red DIALYZER connector from the shunt door
and HOLD IT UP as seen in the Figure right!
f) Continuous flow, more than 0.2 ml per minute from the connector?
Yes
Continuous flow! Turn the machine off THEN proceed to
page 527
18A – DIAGNOSTIC VALVE LEAK TESTS
No
No Flow! See parts a AND b below:
a)
Return the connector to the shunt and close the door!
b) From debug screen 0, Figure right, watch valve #41’s ‘dot’
for thirty (30) seconds. TWO (2) possible scenarios:
1)
IF (and ONLY if) valve #41’s ‘dot’ cycles between white and blue:
Turn
the machine off THEN proceed to
page 527
2)
IF valve #41’s ‘dot’ STAYS white!
Call debug screen 4. Once again, watch
PDIA
AND
ADIA
for one (1) minute!
TWO (2) possible scenarios i) or ii)
below:
i)
IF (and ONLY if) PDIA AND / OR ADIA is NOT between 2.0 and 7.5
OR changes more than 0.1 i.e. is unstable:
Proceed to
page
, procedure number
ii)
IF PDIA AND ADIA is between 2.0 and 7.5 AND does NOT change
more than 0.1:
Proceed to
page 53
, procedure number
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...