2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 324
Scenario #1:
Referring to Figure A (below), approximately 30 ml pulses that stop completely
between each and every
cycle i.e. Pulse → Stop → Pulse → Stop:
See parts a
AND b below
:
a) Reconnect the drain.
b) ENSURING [Dialysate Flow] is ON and set to 800 ml/min, see procedure
(page 325).
Figure A
Scenario #2:
Referring to
Figure B (below), a
pproximately 30 ml Pulse → Stop → Pulse →
‘Dribble’ (a noticeably weaker stream that last about two (2) seconds
) → etc.
Perform parts a and b below:
a) A procedure in another section is performed next. NOTE this page number as
you may be prompted to return here!
b) BEFORE continuing to part c proceed to page
TESTING FOR A LEAKING BALANCING CHAMBER DIAPHRAGM
c) If a leaking Balancing Chamber diaphragm was not located in part b, return to
Dialysis Program (“Select Program” → ‘Dialysis’ → ‘CONFIRM’)!.
d) Allow four (4) minutes then see procedure number
(page 325).
.
Figure B
LEFT BLANK INTENTIONALLY