![Fresenius Medical Care 2008K2 Скачать руководство пользователя страница 136](http://html.mh-extra.com/html/fresenius-medical-care/2008k2/2008k2_troubleshooting-manual_2331635136.webp)
2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 132
MOTORS- 3.1.0 ISOLATE A POTENTIAL ‘BINDING’ MOTOR
a) Return the lines to the shunt!
b)
IMPORTANT!
To prevent damage turn the machine OFF!
c) Figure right, remove the pump head from the
NOTED
motor to
expose the ‘Drive Magnet’.
d) Manually spin the ‘Drive Magnet’! Does it rotate freely i.e. is NOT
binding?
Yes
Rotates freely!
A)
Ensure rotating the drive magnet causes the motor shaft to rotate;
B)
If
the shaft rotates a problem is not indicated at this time but the motor may be stalling
intermittently. Consider replacing the brushes!
No
DOES NOT rotate freely! The motor is bad
*
and brushes WON’T help
.
*
The binding motor may have destroyed the Actuator-Test and/or the Power Logic Board
LEFT BLANK INTENTIONALLY
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...