
2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 602
P- F.1.0 TROUBLESHOOT SCREEN PROBLEM
a) Turn the machine OFF!
b) Per the Figure below, ENSURE the aluminum-looking Data Display Cable (P/N 190194), between the
functional board and the screen, is plugged in CORRECTLY at BOTH sides.
c) Turn the machine on and allow fifty (50) seconds. If the screen continues to display ‘weird’ or
NOTHING continue to part d.
d) Turn the machine OFF!
e) Swap in a known good Data Display Cable (P/N 190194).
f) Turn the machine on and allow fifty (50) seconds.
g) If the screen continues to display ‘weird’ or NOTHING continue to part h.
h) Turn the machine OFF!
i) Swap in a known good Functional Board.
j) Turn the machine on and allow fifty (50) seconds.
NOTE:
A “Cond Offset” banner may appear due to the
different Functional board. Ignore it for now. If the screen continues to display ‘weird’ or NOTHING the
screen may be bad.
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...