2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 102
F- 9.0.6 ACFS = 0
a) Return CFS #10’s connector to distribution board position “x10, CFS”. If returned properly
ACFS
=
between 3.0 and 8.1.
b) From the Home screen, ENSURE [Dialysate Flow] is set to
500 ml/min
.
c) Is the TMP window RED?
Yes
Red TMP window!
RESET SEQUENCE
:
A)
Press and release the ‘Reset’ key then
immediately press and hold it for three (3) seconds;
B)
Allow thirty (30) seconds.
C)
If a
TMP alarm reoccurs attempt the RESET SEQUENCE up to twice more BEFORE
continuing to procedure number F- 9.0.7 (page 102).
No
TMP window is pale yellow/white! See procedure number F- 9.0.7 (page 102).
F- 9.0.7 ANALYZE TMP
TMP is STABLE if (and ONLY if) the TMP Window REMAINS white AND the TMP does NOT change more
than +/- 60 mmHg in three (3) minutes. TWO (2) possible scenarios:
1)
IF (and ONLY if) TMP is STABLE:
(page 102).
2)
IF TMP is UNSTABLE:
A procedure, in a different Section, is performed next.
IMPORTANT! NOTE
this page and procedure number (F- 9.0.7) as you may prompted to return to here then perform
parts A and B below:
A) BEFORE continuing to part B, proceed to
page 527
SECTION 18A – DIAGNOSTIC VALVE
B) If a balancing chamber valve leak was not located in part A, see procedure number
102).
F- 9.0.8 BALANCING CHAMBER DIAPHRAGM TEST
A procedure, in a different Section, is performed next.
IMPORTANT!
NOTE
this page and procedure
number (F- 9.0.8) because you may prompted to return to here:
a) BEFORE continuing to part b, proceed to
page 535
, to perform
LEAKING BALANCING CHAMBER DIAPHRAGM
b) If a leaking diaphragm is not located in part a, return to Dialysis Program
(“Select Program” → ‘Dialysis’
‘CONFIRM’)!
c) From the Home screen, set [Dialysate Flow] to 800 ml/min and press ‘CONFIRM’.
d) If
BC Switch
(debug screen 6) = 897 (constantly) AND
ACFS
(debug screen 10) remains between 3.0
and 8.0. TWO (2) possible bad components:
1)
CFS transducer #10
a
OR;
2)
Sensor board
b,c
.
a
To LOCATE CFS transducer #10 refer to
(page 92);
b
To LOCATE the Sensor board see
c
To prevent “Cond Offset Failure” place the machine into
T and C Mode
(refer to
OPERATING
MODES
page
Error! Bookmark not defined.
)).
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...