2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 609
P- G.0.0 ISOLATE POWER LOGIC BOARD
a)
IMPORTANT! To prevent damage turn the machine OFF!
b) Figure below, trace the cable from the Blood Pressure Module to ENSURE it is NOT reverse connected
with another module! If it first correct this problem THEN re-m5 and +12 volts as this may be
the problem!
c) See procedure number P- G.2.0 (page 609).
P- G.2.0 MODULES OKAY / ISOLATE POWER LOGIC BOARD
a)
Swap in a known good* Power Logic Board.
The board
is known good if +5 and +12 volts
are present in another machine.
*
To LOCATE the board see
(page 9)).!
b)
IMPORTANT! Turn the machine on (fan running)!
c)
ENSURE the meter’s black lead remains connected to chassis ground!
d) Re-m12 and / or +5 volts (within range now?). If the symptom was intermittent call debug
screen 1 to watch
5V Est
and / or
12V Est.
Allowing sufficient time, does the symptom you were
troubleshooting reoccur?
Yes
Symptom reoccurs! See procedure number
(page 610).
No Screen on and / or good voltage(s) now! The previous Power Logic Board may be bad.
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...