2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 101
F- 9.0.5 GOOD FLOW PUMP PRESSURE / ISOLATE CFS CIRCUIT
a) Return to Dialysis Program (“Select Program”
→ ‘Dialysis’ → ‘CONFIRM’)!
b) From the Home screen, set [Dialysate Flow] to 500 ml/min and press ‘CONFIRM’.
c) Figure below, insert one of the resistor plugs, from the
FOUR-RESISTOR SET
, into CFS #10’s
distribution board position, “x10, CFS”.
d) Call debug screen 10.
ACFS
should = 0.0?
Yes
ACFS
= 0.0! See procedure number F- 9.0.6 (page 102).
No
ACFS
DOES NOT = 0.0! See parts a AND b below:
a) ENSURE the resistor plug was placed correctly at position “CFS” before continuing! If
not, repeat procedure number F- 9.0.5 (page 101) from part c.
b)
Leaving the resistor plug installed for now, FOUR (4) possible bad components:
1)
Actuator-Test Board
1
OR;
2)
Sensor Board
2
OR
3)
Sensor Board cable
3
OR;
4)
Distribution board.
1
A)
With the machine off,
swap in a known good Actuator-Test Board
a
;
B)
Return to
Dialysis Program (
“Select Program” → Dialysis’ → ‘CONFIRM’);
C)
If
ACFS
= 0.0 the
previous Actuator-Test Board is bad.
a
To LOCATE the board refer to Figure 4A (page 9)
2
A)
With the machine off, swap in a known good Sensor Board
b
;
B)
Place the machine
into
T and C Mode
(refer to OPERATING MODES (page
Error! Bookmark not
defined.
);
C)
IMPORTANT!
Return to Dialysis Program. If
ACFS
= 0.0 the previous
Sensor Board is bad.
b
To LOCATE the board refer to Figure 4A (page 9)
3
The Sensor Board cable can be checked.
NOTE
three (3)
CFS TRANSDUCER
connections will be checked and proceed to
page 524
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...