![Fresenius Medical Care 2008K2 Скачать руководство пользователя страница 62](http://html.mh-extra.com/html/fresenius-medical-care/2008k2/2008k2_troubleshooting-manual_2331635062.webp)
2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 58
F- 6.2.0 TROUBLESHOOT ‘CONTINUOUS’ DRAIN FLOW
Drain flow NEVER stops! From the Home screen, is [
Conductivity
] more than 15.0 mS?
Yes
[
Conductivity
] more than 15.0! ENSURING Loading Pressure (Rinse port gauge) is
PEAKING to between 23 and 27 psi, see procedure number
(page 59).
No
[
Conductivity
] remaining less 15.0! This procedure checks if Valve #43 is sticking open.
See parts a AND b below:
a)
Per the Figure below
, double clamp valve #43’s OUTPUT tubing.
NOTE:
Valve
#43’s output tubing extends towards the front of the machine!
b) Recheck drain flow. TWO (2) possible scenarios:
1)
IF (and ONLY if) drain flow becomes
Pulse → Stop → Pulse → Stop
:
Remove the clamps from valve #43. TWO (2) possible bad components:
1)
Bad
Actuator-Test
*
Board OR;
2
) Bad valve #43.
*
To LOCATE the Actuator-Test board, refer to
(page 9)
2)
IF drain flow continues to be ‘continuous’:
Remove the clamps from valve
#43 then proceed to
page 61
, procedure number
Figure 11 – Valve #43 Location
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...