2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 416
FIL- 2.0.0 VERIFY DIALYSATE SETTINGS
a)
IMPORTANT! CLOSE THE SHUNT DOOR!
b) Is the external flow indicator’s ‘bob’ moving up and down in the sight tube?
Yes
‘Bob’ moving! See procedure number
(page 416).
No a) At the bottom of the screen, select the
‘Dialysate’ tab.
b) Figure right, if necessary adjust the
Conductivity Limits until ‘Actual’ Conductivity is
CENTERED between them.
c)
IMPORTANT!
Press ‘CONFIRM’.
d) Allow up to three (3) minutes for the
Conductivity window to turn white (i.e. No
Cond alarm).
e) Call debug screen 0. Figure right, allow Valve #24’s
‘dot’ to turn BLUE BEFORE continuing to part f!
f) Is the flow indicator’s ‘bob’ moving up and down now?
Yes
‘Bob’ moving!’ See procedure number
(page 416).
No ‘Bob’ NOT moving!
ENSURING:
1)
From
the Home screen, [Dialysate Flow] is set to
800 ml/min! AND;
2)
Screen 0’s Valve #24’s ‘dot’ is blue! If (and ONLY if) ‘bob’ is still
not moving proceed to
page 20
SECTION 1 - FLOW ERRORS IN DIALYSIS
FIL- 2.0.1 ‘BOB’ MOVING / ISOLATE POSSIBLE UF PUMP LEAK
a) Figure right, ENSURE the lamp above the front panel’s UF on/off key
is
OFF
i.e. UF is OFF!
Parts b and c next page
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...